8.1.8. Relay Outputs (Optional) Page Default Values


Press

Select

to choose relay. If

Relay 0

displays, no relays are installed.

Menu Item

Display

Default

Range

Type of relay

Relay 1 RelayType

Off

Off, Alarm, Cycl, Ev1 to Ev4

Action of relay

Relay 1 RelayAction

NDE

NDE, NE, NDEL, NEL

26

Type of alarm

27

Relay 1 AlarmType

High

High, Low, Plus, Minus, Band

Alarm setpoint

27

Relay 1 AlmHi/Lo SP

538°C

-73°C to 760°C (-100°F to 1400°F)

Alarm deviation

27

Relay 1 AlmDevBand

3°C

1 to 56°C (1 to 100°F)

Inhibit alarm

27

Relay 1 ALMInhibit

On

En or Dis

Type of event

28

Relay 1 EventType

Time

Time or Temp

Event setpoint

29

Relay 1 Event SP

SPUpLim

SPLoLim to SPUpLim

30

(repeat for relay outputs 2-4, if available)


26

Normally de-energized and non-latching, normally energized and non-latching, normally de-energized

and latching, normally energized and latching.

27

Appears only for alarm types.

28

Appears only for time or temperature event types.

29

Appears only for temperature event types.

30

See enable page.