8.1.8. Relay Outputs (Optional) Page Default Values
Press
Select
to choose relay. If
Relay 0
displays, no relays are installed.
Menu Item
Display
Default
Range
Type of relay
Relay 1 RelayType
Off
Off, Alarm, Cycl, Ev1 to Ev4
Action of relay
Relay 1 RelayAction
NDE
NDE, NE, NDEL, NEL
Type of alarm
Relay 1 AlarmType
High
High, Low, Plus, Minus, Band
Alarm setpoint
Relay 1 AlmHi/Lo SP
538°C
-73°C to 760°C (-100°F to 1400°F)
Alarm deviation
Relay 1 AlmDevBand
3°C
1 to 56°C (1 to 100°F)
Inhibit alarm
Relay 1 ALMInhibit
On
En or Dis
Type of event
Relay 1 EventType
Time
Time or Temp
Event setpoint
Relay 1 Event SP
SPUpLim
SPLoLim to SPUpLim
(repeat for relay outputs 2-4, if available)
26
Normally de-energized and non-latching, normally energized and non-latching, normally de-energized
and latching, normally energized and latching.
27
Appears only for alarm types.
28
Appears only for time or temperature event types.
29
Appears only for temperature event types.
30
See enable page.