Manual mode – Despatch Protocol Plus Controller (Heat Only) User Manual
Page 22
O
PERATION
Protocol Plus Owner’s Manual
22
Version 2
Copyright © 2010 by Despatch Industries.
All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.
O
Control can be configured to automatically activate Manual,
Timer or Profile Mode when power is applied (power switch ON).
Refer to Section 5.5.
5.2. Manual Mode
1. Press
Select
until
Manual
displays.
O
Press Run at any time in the procedure to activate Manual Mode.
2. Press
Menu
to display the Process Temperature Setpoint (
Setpt
). Change Setpoint
(
Setpt
) by pressing
▲
or
▼
.
O
If the SPChange on the Enable page in Setup Mode has been set
to DISABLED, it must be changed to ENABLED before any
changes to the process temperature and high limit setpoints can
be made.
3. Press
Menu
a second time to display current high limit setpoint (
HLSP
). The value
can be adjusted pressing
▲
or
▼
. If
Band
displays, the High Limit band feature is
activated (see Control page, Section 5.6.7) and the high limit cannot be adjusted.
4. Optional feature: Press
Menu
a third time to display Event1. Press
▲
to turn the
event ON or
▼
to turn the even OFF. Repeat for all events which are enabled (up to
four events).
5. To start Manual Mode, press
Run
.
The display changes from
Stop
to
Run
. Return to Stopped Mode by pressing
Stop
. While
in operation, the process setpoint can be adjusted by using
▲
or
▼
to change the value
while the mode is running. Press
Menu
to display the High Limit Setpoint (HLSP)
setting.
O
Changes to High Limit Setpoint or Event Output Configuration
must be accomplished from the Stop Mode.