Protocol Plus Owner’s Manual

O

PERATION

Version 2

39

Copyright © 2010 by Despatch Industries.

All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.

certain conditions where the center of the oven chamber is not the same temperature as
the control thermocouple. This temperature differential may occur when the oven is not
allowed to soak at a constant temperature for long periods of time, or when the oven is
being used at high temperature. Table 11 shows the menu items available.

5.6.10.3. Access Zone Calibration Page

1. Enter Setup Mode.
2. Press

Page

until

Zone Cal

displays.

3. Press

Menu

.

4. Change each parameter by pressing

Menu

until the desired parameter is displayed,

and then press

▲ or ▼ to change the value.

5. Press

Page

to exit the Zone Calibration Page.

O


Use Factory Calibration only when no calibration instrument is
available. Push

▲ to restore factory values.

Table 11. Zone Calibration Page Parameters and Values.
Menu Item

Display

Description

Range

Zone 1
actual

Zone Cal
Zone1Act

Point at which Zone 1 is set
(center of chamber)

-73°C to 760°C (-100°F to
1400°F)

Zone1
displayed

Zone Cal
Zone1Dis

Desired displayed value for Zone
1 setting

-73°C to 760°C (-100°F to
1400°F)

Zone 2
actual

Zone Cal
Zone2Act

Point at which Zone 2 is set
(center of chamber)

-73°C to 760°C (-100°F to
1400°F)

Zone2
displayed

Zone Cal
Zone2Dis

Desired displayed value for Zone
2 setting

-73°C to 760°C (-100°F to
1400°F)

Factory
calibration

15

Zone Cal
FactCal

Restores the factory calibration
values

Ready or Done (push-key)



5.6.10.4. Access Factory Calibration Recovery

Factory Calibration Recovery restores the original factory calibration values. This may be
helpful if the calibration has been lost and a calibration instrument is not readily
available. To use the Factory Calibration Recovery feature only, bypass the Zone 1 and
Zone 2 calibration parameters by pressing

Menu

.

15

Use only when no calibration instrument is available, push

▲ to restore factory values.