beautypg.com

2 technical data, Battery, Power unit – Eppendorf Research pro User Manual

Page 5: 2 technical data 2, Technical data

background image

5

The technical data is valid only when the Research pro is used with Eppendorf pipette tips. The
multi-channel data is valid for eight- and twelve-channel pipettes. The 50 – 1,200 model is
available as an eight-channel version only.

Test conditions in accordance with ISO 8655 for piston-stroke pipettes with an air cushion by
means of a standardized fine balance with a moisture trap. Number of determinations:
10 pipettings; degassed distilled water, 20 °C – 25 °C ± 0.5 °C; maximum speed; PIP in
standard mode of operation (Sec. 5.3); prewetted pipette tip; dispensing onto inside of tube.
If the place where the pipette is used is at extremely high altitude, an adjustment must be made
in line with the ambient air pressure.

Battery
Nickel-metal hydride battery pack, 1,200 mAh / 2.4 V reversible overcurrent and over-
temperature protection. Overcharging protection when connected to Research pro.
Charging time: approx. 9.5 hours for a fully discharged battery.

Power unit
Input voltage, country-specific: 230 V ± 10 %, 50/60 Hz; 120 V ± 10 %, 60 Hz;
100 V ± 10 %, 50/60 Hz; 240 V ± 10 %, 50 Hz
Output voltage: 9 V DC; 200 mA (1.8 VA)

Technical specifications subject to change!

Single-channel

Volume range Step size

Color

Pipette tip

Volume

Max. permissible error

code

epT.I.P.S.

Error

systematic

random

µL

µL

µL

%

µL

%

µL

0.5 –

10

0.01

light
gray

20 µL L

1
5

10

± 2.5
± 1.5
± 1.0

± 0.025
± 0.075
± 0.1

± 1.8
± 0.8
± 0.4

± 0.018
± 0.04
± 0.04

5 –

100

0.1

yellow

200 µL

10
50

100

± 2.0
± 1.0
± 0.8

± 0.2
± 0.5
± 0.8

± 1.0
± 0.3
± 0.2

± 0.1
± 0.15
± 0.2

20 –

300

0.5

orange

300 µL

30

150
300

± 2.5
± 1.0
± 0.6

± 0.75
± 1.5
± 1.8

± 0.7
± 0.3
± 0.2

± 0.21
± 0.45
± 0.6

50 – 1,000

1

blue

1,000 µL

100
500

1,000

± 3.0
± 1.0
± 0.6

± 3
± 5
± 6

± 0.6
± 0.2
± 0.2

± 0.6
± 1
± 2

100 – 5,000

10

violet

5 mL

500

2,500
5,000

± 3.0
± 1.2
± 0.6

± 15
± 30
± 30

± 0.6
± 0.25
± 0.15

± 3
± 6.25
± 7.5

Multi-channel

0.5 –

10

0.01

light
gray

20 µL L

1
5

10

± 5.0
± 3.0
± 2.0

± 0.05
± 0.15
± 0.2

± 3.0
± 1.5

± 0.8

± 0.03
± 0.075
± 0.08

5 – 100

0.1

yellow

200 µL

10
50

100

± 2.0
± 1.0
± 0.8

± 0.2
± 0.5
± 0.8

± 2.0

± 0.8

± 0.25

± 0.2
± 0.4
± 0.25

20 – 300

0.5

orange

300 µL

30

150
300

± 2.5
± 1.0
± 0.6

± 0.75
± 1.5
± 1.8

± 1.0

± 0.5
± 0.25

± 0.3
± 0.75
± 0.75

50 – 1,200

5

green

1,250 µL

120
600

1,200

± 6.0
± 2.7
± 1.2

± 7.2

± 16.2
± 14.4

± 0.9
± 0.4
± 0.3

± 1.08
± 2.4
± 3.6

Technical data

2 Technical data

2

E01 bis 04 Seite 5 Donnerstag, 24. September 2009 3:04 15