2 technical data, Battery, Power unit – Eppendorf Research pro User Manual
Page 5: 2 technical data 2, Technical data
![background image](/manuals/551344/5/background.png)
5
The technical data is valid only when the Research pro is used with Eppendorf pipette tips. The
multi-channel data is valid for eight- and twelve-channel pipettes. The 50 – 1,200 model is
available as an eight-channel version only.
Test conditions in accordance with ISO 8655 for piston-stroke pipettes with an air cushion by
means of a standardized fine balance with a moisture trap. Number of determinations:
10 pipettings; degassed distilled water, 20 °C – 25 °C ± 0.5 °C; maximum speed; PIP in
standard mode of operation (Sec. 5.3); prewetted pipette tip; dispensing onto inside of tube.
If the place where the pipette is used is at extremely high altitude, an adjustment must be made
in line with the ambient air pressure.
Battery
Nickel-metal hydride battery pack, 1,200 mAh / 2.4 V reversible overcurrent and over-
temperature protection. Overcharging protection when connected to Research pro.
Charging time: approx. 9.5 hours for a fully discharged battery.
Power unit
Input voltage, country-specific: 230 V ± 10 %, 50/60 Hz; 120 V ± 10 %, 60 Hz;
100 V ± 10 %, 50/60 Hz; 240 V ± 10 %, 50 Hz
Output voltage: 9 V DC; 200 mA (1.8 VA)
Technical specifications subject to change!
Single-channel
Volume range Step size
Color
Pipette tip
Volume
Max. permissible error
code
epT.I.P.S.
Error
systematic
random
µL
µL
µL
%
µL
%
µL
0.5 –
10
0.01
light
gray
20 µL L
1
5
10
± 2.5
± 1.5
± 1.0
± 0.025
± 0.075
± 0.1
± 1.8
± 0.8
± 0.4
± 0.018
± 0.04
± 0.04
5 –
100
0.1
yellow
200 µL
10
50
100
± 2.0
± 1.0
± 0.8
± 0.2
± 0.5
± 0.8
± 1.0
± 0.3
± 0.2
± 0.1
± 0.15
± 0.2
20 –
300
0.5
orange
300 µL
30
150
300
± 2.5
± 1.0
± 0.6
± 0.75
± 1.5
± 1.8
± 0.7
± 0.3
± 0.2
± 0.21
± 0.45
± 0.6
50 – 1,000
1
blue
1,000 µL
100
500
1,000
± 3.0
± 1.0
± 0.6
± 3
± 5
± 6
± 0.6
± 0.2
± 0.2
± 0.6
± 1
± 2
100 – 5,000
10
violet
5 mL
500
2,500
5,000
± 3.0
± 1.2
± 0.6
± 15
± 30
± 30
± 0.6
± 0.25
± 0.15
± 3
± 6.25
± 7.5
Multi-channel
0.5 –
10
0.01
light
gray
20 µL L
1
5
10
± 5.0
± 3.0
± 2.0
± 0.05
± 0.15
± 0.2
± 3.0
± 1.5
± 0.8
± 0.03
± 0.075
± 0.08
5 – 100
0.1
yellow
200 µL
10
50
100
± 2.0
± 1.0
± 0.8
± 0.2
± 0.5
± 0.8
± 2.0
± 0.8
± 0.25
± 0.2
± 0.4
± 0.25
20 – 300
0.5
orange
300 µL
30
150
300
± 2.5
± 1.0
± 0.6
± 0.75
± 1.5
± 1.8
± 1.0
± 0.5
± 0.25
± 0.3
± 0.75
± 0.75
50 – 1,200
5
green
1,250 µL
120
600
1,200
± 6.0
± 2.7
± 1.2
± 7.2
± 16.2
± 14.4
± 0.9
± 0.4
± 0.3
± 1.08
± 2.4
± 3.6
Technical data
2 Technical data
2