5 pipetting with separate rinsing (rinse = rns), 5 operation 5, Operation – Eppendorf Research pro User Manual
Page 13

13
Rinsing (RNS) is recommended for volumes below 10 µL. RNS is suitable for mixing the
dispensing volume and the specimen liquid when large volumes are used.
It is strongly recommended to use the "max. speed" setting.
Programming
Procedure
If PROG appears in
the display: Press
the
key
briefly
until PROG
disappears.
Select "PIP" for pipetting
.
Volume selection (here: 5 µL).
The volume rocker has two
pressure points for fast or slow
volume adjustment for up and
down.
If necessary, press the Option
key several times until RNS
appears in the display. RNS is
selected for rinsing.
If
appears in the display: Press the
key briefly.
Liquid is aspirated.
Rinsing/mixing takes place three
times after liquid has been
dispensed.
Mixing volume results from
dispensing volume and the
volume of the blow-out step
(BLO).
If the Actuate key is held down
when liquid is dispensed,
mixing takes place until the key
is released.
Return to basic position by
pressing the key when device
is not immersed in the liquid.
The pipette is now once again
ready to aspirate.
5.5 Pipetting with separate rinsing (Rinse = RNS)
Select
Program
Volume
Option
Option
Option
R
R
Alternative mixing procedure
Operation
5 Operation
5