beautypg.com

5 pipetting with separate rinsing (rinse = rns), 5 operation 5, Operation – Eppendorf Research pro User Manual

Page 13

background image

13

Rinsing (RNS) is recommended for volumes below 10 µL. RNS is suitable for mixing the
dispensing volume and the specimen liquid when large volumes are used.
It is strongly recommended to use the "max. speed" setting.

Programming

Procedure

If PROG appears in
the display:
Press
the

key

briefly

until PROG

disappears.

Select "PIP" for pipetting

.

Volume selection (here: 5 µL).
The volume rocker has two
pressure points for fast or slow
volume adjustment for up and
down.

If necessary, press the Option
key several times until RNS
appears in the display. RNS is
selected for rinsing.

If

appears in the display: Press the

key briefly.

Liquid is aspirated.

Rinsing/mixing takes place three
times after liquid has been
dispensed.
Mixing volume results from
dispensing volume and the
volume of the blow-out step
(BLO).

If the Actuate key is held down
when liquid is dispensed,
mixing takes place until the key
is released.

Return to basic position by
pressing the key when device
is not immersed in the liquid.

The pipette is now once again
ready to aspirate.

5.5 Pipetting with separate rinsing (Rinse = RNS)

Select

Program

Volume

Option

Option

Option

R

R

Alternative mixing procedure

Operation

5 Operation

5

E05_1 Bedienung A5 Seite 13 Dienstag, 19. Dezember 2006 2:05 14