4 pipetting with separate blow-out (blow = blo), 5 operation, Operation – Eppendorf Research pro User Manual
Page 12

12
BLO is recommended for use with liquids with a high wetting power or with liquids which are
prone to the formation of foam (e.g. solutions containing protein). The residual liquid is
dispensed by pressing the Actuate key separately.
Programming
Procedure
If PROG appears in
the display: Press
the
key
briefly
until PROG
disappears.
Select "PIP" for pipetting
.
Volume selection (here:
800 µL). The volume rocker has
two pressure points for fast or
slow volume adjustment for up
and down.
If necessary, press the Option
key several times until BLO
appears in the display. BLO is
selected for blow-out.
If
appears in the display: Press the
key briefly.
Liquid is aspirated.
Liquid is dispensed.
Blow-out:
Delayed dispensing of residual
liquid.
If the Actuate key is held down
during the blow-out, the
return does not take place
until the key has been
released.
The pipette is now once again
ready to aspirate.
5.4 Pipetting with separate blow-out
(Blow = BLO)
Select
Program
Volume
Option
Option
Option
R
R
Alternative blow-out procedure
➞
5 Operation
Operation
5