beautypg.com

7 troubleshooting 7, Troubleshooting – Eppendorf Research pro User Manual

Page 31

background image

31

Error

Cause

Solution

Pipette is dripping;
the volume
dispensed is
incorrect.

The tip may be loose.

There may be foreign
matter between the
pipette and the tip.

The piston, nose cone and
cylinder may be damaged
or contaminated.

The lower part may be
loose.

For multi-channel model:
The O-ring may be
damaged.

Attach the tip firmly. Make sure that
an Eppendorf tip is used.

Wipe the pipette. Ensure that the tip
is protected from dust.

Treat the lower half of the pipette as
described in Sec. 6 and Part B
"Ordering Information/service
parts".

Residual liquid is
in the tip;
dispensing is
incomplete.

See above.

The dispensing speed
may be too high.

The incorrect pipetting
option may have been
selected.

See above.

The procedure for checking the
function is described in Sec. 5,
"Operation".

Motor stops
during the
dispensing
procedure.

The battery may be run
down.

The pipette may be
heavily contaminated.

Discard the dispensing as it is
incorrect. Recharge the battery
(Sec. 3.3) or treat the pipette as
described in Sec. 6.

Multi-channel:
Pipette doesn’t
aspirate liquid.

(Continued
on next page)

The magnetic coupling on the
multi-channel lower part no
longer works.

In the case of the Research
pro Pipette 4860, strong
impact stress can result in a
separation of the magnetic
connection between the
spindle and the piston
actuator. As a result of this
separation, the pipette no
longer aspirates liquid, despite
movement of the motor.

In order to reestablish the connection,
proceed as follows:

– Separate the ejector from the lower

part.

– To do this, press the lower part

against a surface until it springs in.
Simultaneously press the stop clutch
inward with side "B" of the universal
tool. The ejector cover panel springs
up.

Troubleshooting

7 Troubleshooting

7

E06 Pflege etc. Seite 31 Dienstag, 19. Dezember 2006 2:16 14