beautypg.com

5 operation 5, Operation – Eppendorf Research pro User Manual

Page 23

background image

23

Programming

Programming can be ended prematurely and then stored by pressing the

key. EDIT

disappears from the display.
If EDIT is ended with the

key before completion of programming for the memory slot, the

changes are not transferred (emergency exit).

Procedure

The procedure depends on the pipetting option selected.
Depending on the option, procedure is identical to Sections 5.3 to 5.7.

When Prog. 1 – 5 are carried out, the Option and Speed keys as well as the Volume rocker are
all disabled.

Double-click

the key until

PROG

appears in the display.

Select the memory slot.

Hold down

the key until

EDIT

appears.

The input field flashes. Select

PIP

.

Press the Actuate key.

The input field flashes. Select
the volume and store.

Select the aspirating and
dispensing speed, and store.
The direction arrow flashes.

The option is selected and
stored.
(The magnifying glass shows
the standard mode.)

Programming is now complete
(here: standard mode; no
option).
EDIT disappears.

5.10.1 Example for programming pipetting in the program level

Program

12 3 4 5

PROG

Select

PROG

12 3 4 5

Program

EDIT

1

PROG 2 3 4 5

EDIT

Select

EDIT

1

PROG 2 3 4 5

ENTER

Volume

or

Select

EDIT

1

PROG 2 3 4 5

ENTER

Select

EDIT

1

PROG 2 3 4 5

ENTER

Select

EDIT

1

PROG 2 3 4 5

ENTER

1

PROG 2 3 4 5

Program

R

Operation

5 Operation

5

E05_2 Bedienung A5 Seite 23 Dienstag, 19. Dezember 2006 2:07 14