10 Technical data
BioPhotometer plus — Operating manual
40
10
Technical data
10.1 Power supply
10.2 Ambient conditions
Not tropicalized.
Protect against direct sunlight.
10.3 Weight / dimensions
10.4 Interfaces
The printer to be connected must meet the requirements of EN 60950 or UL 1950.
Power supply
100 to 240 V ± 10 % / 50 to 60 Hz ± 5 %
Overvoltage category
IEC 61010-1 category II
Degree of contamination:
IEC 61010-1 category II
Power consumption
approx. 20 W during operation,
approx. 10 W in standby mode
Permitted mains interruption
approx. 10 ms at 90 V
approx. 20 ms at 220 V
Fuses
T 1A/250 V, 5 mm x 20 mm (2 off)
Operation
Ambient temperature: 15 to 35 °C
Rel. humidity: 15 to 70%
Atmospheric pressure: 86 to 106 kPa
Storage
Ambient temperature: -25 to 70 °C
Rel. humidity: 15 to 70%
Atmospheric pressure: 30 to 106 kPa
Weight
3 kg (packaged: 4.8 kg)
Dimensions
Width: 200 mm (packaged: 290 mm)
Depth: 320 mm (packaged: 430 mm)
Height: 100 mm (packaged: 200 mm)
Space required
Width: 400 mm (or 650 mm including printer)
Depth: 500 mm
Interface for printer and PC
serial RS-232