beautypg.com

Technical data, 1 power supply, 2 ambient conditions – Eppendorf BioPhotometer plus User Manual

Page 40: 3 weight / dimensions, 4 interfaces, 10 technical data, Power supply, Ambient conditions, Weight / dimensions, Interfaces

background image

10 Technical data

BioPhotometer plus — Operating manual

40

10

Technical data

10.1 Power supply

10.2 Ambient conditions

Not tropicalized.
Protect against direct sunlight.

10.3 Weight / dimensions

10.4 Interfaces

The printer to be connected must meet the requirements of EN 60950 or UL 1950.

Power supply

100 to 240 V ± 10 % / 50 to 60 Hz ± 5 %

Overvoltage category

IEC 61010-1 category II

Degree of contamination:

IEC 61010-1 category II

Power consumption

approx. 20 W during operation, 
approx. 10 W in standby mode

Permitted mains interruption

approx. 10 ms at 90 V
approx. 20 ms at 220 V

Fuses

T 1A/250 V, 5 mm x 20 mm (2 off)

Operation

Ambient temperature: 15 to 35 °C
Rel. humidity: 15 to 70%
Atmospheric pressure: 86 to 106 kPa

Storage

Ambient temperature: -25 to 70 °C
Rel. humidity: 15 to 70%
Atmospheric pressure: 30 to 106 kPa

Weight

3 kg (packaged: 4.8 kg)

Dimensions

Width: 200 mm (packaged: 290 mm)
Depth: 320 mm (packaged: 430 mm)
Height: 100 mm (packaged: 200 mm)

Space required

Width: 400 mm (or 650 mm including printer)
Depth: 500 mm

Interface for printer and PC

serial RS-232