Troubleshooting, 1 result flags, Troubleshooting 8.1 – Eppendorf BioPhotometer plus User Manual

Page 36: Result flags, 8 troubleshooting, Oper at ing man u al

background image

8 Troubleshooting

BioPhotometer plus — Operating manual

36

8

Troubleshooting

8.1

Result flags

Result flags

Possible cause

Remedy

0.015 A

340

Only for nucleic acid methods and 
"PROTEIN 280 nm":
The marking '¶' of the A

340

value indicates that

the absorbance values for 230, 260 and 280
nm are corrected with the absorbance for 340
nm (parameter "CORRECTION A

340

" is

enabled).

None. The flag is only for information.

0.015 A

550

or
0.015 A

650

Only for method groups "Dye 550" and 
"Dye 650":
The marking '¶' of the A

550

value or the A

650

value indicates that the absorbance values for
260 and 280 nm are corrected with the
absorbance for 550 or 650 nm (parameter
"CORRECTION

550 (or 650)

" is enabled).

None. The flag is only for information.

++++++

The measured absorbance is above A = 3.

Dilute sample.

Check cuvette volume (light path
height is 8.5 mm).

Clean cuvette shaft (see Cleaning
on page 32) a
nd insert cuvette the
right way round: measuring
surface towards the light path. The
direction of the light path is
indicated on the blue cuvette shaft
cover by an arrow.

Use cuvette from a material which
is optically transparent for light of
the measuring wavelength (for
measurements in UV e.g.:
Eppendorf UVette).

------

(Instead of a value for the ratio:)
Ratio cannot be calculated because one of the
absorbance values used for the calculation is
A = 0 or A > 3.

Repeat measurement, if necessary
with a diluted sample.