Troubleshooting, 1 result flags, Troubleshooting 8.1 – Eppendorf BioPhotometer plus User Manual
Page 36: Result flags, 8 troubleshooting, Oper at ing man u al

8 Troubleshooting
BioPhotometer plus — Operating manual
36
8
Troubleshooting
8.1
Result flags
Result flags
Possible cause
Remedy
0.015 A
340
¶
Only for nucleic acid methods and
"PROTEIN 280 nm":
The marking '¶' of the A
340
value indicates that
the absorbance values for 230, 260 and 280
nm are corrected with the absorbance for 340
nm (parameter "CORRECTION A
340
" is
enabled).
None. The flag is only for information.
0.015 A
550
¶
or
0.015 A
650
¶
Only for method groups "Dye 550" and
"Dye 650":
The marking '¶' of the A
550
value or the A
650
value indicates that the absorbance values for
260 and 280 nm are corrected with the
absorbance for 550 or 650 nm (parameter
"CORRECTION
550 (or 650)
" is enabled).
None. The flag is only for information.
++++++
The measured absorbance is above A = 3.
Dilute sample.
Check cuvette volume (light path
height is 8.5 mm).
Clean cuvette shaft (see Cleaning
on page 32) and insert cuvette the
right way round: measuring
surface towards the light path. The
direction of the light path is
indicated on the blue cuvette shaft
cover by an arrow.
Use cuvette from a material which
is optically transparent for light of
the measuring wavelength (for
measurements in UV e.g.:
Eppendorf UVette).
------
(Instead of a value for the ratio:)
Ratio cannot be calculated because one of the
absorbance values used for the calculation is
A = 0 or A > 3.
Repeat measurement, if necessary
with a diluted sample.