3 general errors, General errors, 8 troubleshooting – Eppendorf BioPhotometer plus User Manual
Page 38: Oper at ing man u al

8 Troubleshooting
38
BioPhotometer plus — Operating manual
8.3
General errors
Fault
Possible cause
Remedy
Measuring results are 
imprecise.
•
Shelf life of reagent solution 
exceeded.
Ensure that the reagent is still within its shelf life 
and properly prepared.
•
Reagent not prepared 
correctly.
Use clean demineralized water of adequate 
quality for preparation if required.
•
Pipetting incorrect.
Ensure that the pipette is calibrated and that 
pipetting is being performed correctly.
•
Incubation procedure before 
measurement is incorrect.
If the method procedure requires incubation 
before the measurement, ensure that the 
temperature and time for incubation are correctly 
observed.
•
Cuvette dirty.
Clean and rinse the cuvette. When changing 
cuvette, ensure that the optical surface of the 
cuvette remains clean and has not come into 
contact with fingers.
If the optical surface of the cuvette is soiled with 
fingerprints, clean it by wiping with a lint-free 
laboratory cloth soaked in alcohol.
•
Cuvette not completely full of 
measuring solution without 
bubbles.
Ensure that the necessary minimum volume of 
the cuvette for a measurement is reached and 
there are no bubbles in the measuring solution.
•
Measuring solution turbid.
Centrifuge turbid measuring solutions containing 
particles and use the clear supernatant.
•
Photometer drifting.
Contact Eppendorf Service.
Measuring results incorrect.
•
Method wrongly programmed.
Ensure that the method parameters are entered 
correctly.
•
Standard solution not 
prepared correctly.
Ensure that the correct standard is used and that 
the measuring solution for the standard is 
prepared correctly.
•
Reagent absorbance drifting.
(With unstable reagent absorbance and 
end-point methods): when measuring a long 
series of samples, do not measure the reagent 
blank only at the beginning of the series, but 
also during the series of samples. If the reagent 
blank drifts significantly, the reagent is 
unsuitable for fault-free measurements and must 
be replaced with new reagent.
Other possible causes can be 
found under "Measuring results 
imprecise".
Other possible remedies can be found under 
"Measuring results imprecise".
EN
Oper
at
ing man
u
al
