3 application limits, 4 note on product liability, Application limits – Eppendorf BioPhotometer plus User Manual
Page 12: Note on product liability, 3 safety, Oper at ing man u al
![background image](/manuals/551295/12/background.png)
3 Safety
12
BioPhotometer plus — Operating manual
3.3
Application limits
3.4
Note on product liability
In the following cases, the protection provided in the device may be impaired: Liability for the
function of the device passes to the operator if:
•
the device is not used in accordance with the operating manual.
•
the device is used outside the sphere of application described here.
•
the device is used with accessories and consumables (e.g. tubes and plates), which are not
recommended by Eppendorf AG.
•
the device is maintained or repaired by persons not authorized by Eppendorf.
•
the owner has made unauthorized modifications to the device.
CAUTION!
Caution! Damage from improper cleaning of the cuvette shaft.
Only clean the cuvette shaft using a moist cotton swab.(see Cleaning on page 32)
Do not allow any liquid to enter the cuvette shaft.
Do not reach with your fingers into the cuvette shaft.
CAUTION!
Caution! Faulty measurement due to device confusion.
If you use the devices Biophotometer 6131 and BioPhotometer plus 6132 in your laboratory,
note the different method designations on the keys.
DANGER!
Risk of explosion!
Do not operate the device in rooms where work is being carried out with explosive
substances.
Do not use this device to process any explosive, radioactive or highly reactive substances.
Do not use this device to process any substances, which could create an explosive
atmosphere.