beautypg.com

9 appendix a, Appendix a – Eppendorf FemtoJet express User Manual

Page 52

background image

100

Other error numbers as well as the error cause and solution are described in Section 5,
"Troubleshooting". During remote control, no error texts appear in the FemtoJet

®

express

display.

Error no. during execution
ERR = [eee]

Cause

Solution

070

FemtoJet

®

express is not

in routine mode.

Check FemtoJet

®

express.

Bring FemtoJet

®

express into

routine operation.
Eliminate error.

071

FemtoJet

®

express is not

ready.

Check FemtoJet

®

express or

micromanipulator.
As necessary, wait or eliminate
error.

072

The command is not
finished within the
scheduled time.

Check FemtoJet

®

express or

micromanipulator.
Eliminate error.

073

This command cannot be
executed at the moment,
because of insufficient
storage pressure.
See: Warning 001

Check FemtoJet

®

express.

Eliminate error.

074

The connected micro-
manipulator did not
respond.
See: Warning 003

Check micromanipulator.
Eliminate error.