5 troubleshooting, Troubleshooting – Eppendorf FemtoJet express User Manual
Page 33

81
Warning 5
External pressure is insufficient.
The minimum supply pressure required for an injection or
the Clean function was not present when the function was
triggered.
Possible cause:
There is no external pressure supply, it is not connected or
supplies too little pressure (< 4,000 hPa).
Solution:
Connect external pressure supply and select the external
pressure within the range from 4,000 – 8,000 hPa.
Warning 6
Injection pressure (or compensation pressure) is close top
the external pressure.
An injection pressure is required that is not at least
500 hPa below the external pressure (value displayed to
the upper right).
Possible cause:
External pressure is set too low.
Injection pressure has been (accidentally) set too high.
Injection pressure has been intentionally so set.
Solution:
Increase external pressure or reduce injection pressure.
Note:
This warning is issued only once. Pressing the
key will remove the error message. It will
generally be possible to continue work. However, optimal
valve regulation should not be anticipated.
Warning 7
External pressure is insufficient.
Possible cause:
There is no external pressure supply, it is not connected or
supplies too little pressure (< 4,000 hPa).
Solution:
Connect external pressure supply and select the external
pressure within the range from 4,000 – 8,000 hPa.
WARNING 005
Supply pressure
is too low
INJECT = Continue
WARNING 006
4202
Supply pressure
is lower than +500 hPa
INJECT = Continue
WARNING 007
Supply pressure
is too low
INJECT = Continue