9 appendix a 9, Appendix a – Eppendorf FemtoJet express User Manual
Page 51

99
Command to read out
parameters
[C]
Code
[xxx]
Significance of
output parameter
[A]
Output parameter
[rrrrrrrrrr]
Query program version
101
Versions number
x.xx
Query bleeper
150
Status:
0 = off / 1 = on
0 or 1
Query pressure unit
151
Unit:
0 = hPa / 1 = psi
0 or 1
Query injection time t
i
152
Injection
time
in 1/10 seconds
1 – 999
Query compensation pressure p
c
153
Pressure in hPa /psi
0 – 6,000
/
87
Query injection pressure p
i
154
Pressure in hPa /psi
0 – 6,000
/
87
Query background illumination
155
Status:
0 = off / 1 = on
0 or 1
Error no. for incorrect
operating construction
ERR = [eee]
Cause
Solution
060
The first character is not
a c or C.
Check entry parameters.
Change entry.
061
The 2nd to 4th character
is not a digit.
Check entry parameters.
Change entry.
062
The 5th character is not =.
Check entry parameters.
Change entry.
063
The parameter is not within
the permitted range.
Check entry parameters.
Change entry.
064
The command is not known.
Check entry parameters.
Change entry.
Appendix A
9 Appendix A
9
09_AnhangA_en.fm Seite 99 Freitag, 29. Juni 2007 2:52 14