1 self-test, 5 troubleshooting, Troubleshooting – Eppendorf FemtoJet express User Manual
Page 31

79
5.1.1 Self-test
FUNC 6
The FemtoJet
®
is checked for leaks in the self-test.
This self-test can also be used if a leak is suspected.
The test runs automatically.
The pressure tube must be connected to a closed capillary
(e.g. melted closed or filled) for the pressure control to work
properly.
key: Complete self-test
–
In the case of two different pressures, checks for leaks
in the pressure accumulator (p
s
) and for the internal
tightness of the valves.
–
Checks for leaks in the area of working pressure (p
w
)
in the case of two different pressures.
–
Checks behavior during injection
at different pressures (short continuous test).
If certain values are outside the specified range, an error
message is emitted. If this recurs, inform Service.
key: Self-test to search for leaks
–
Checks for leaks in the pressure accumulator (p
s
) in the
case of two different pressures, as well as for inner
tightness of the valves.
–
Checks for leaks in the area of working pressure (p
w
) in
the case of two different pressures.
If certain values are outside the specified range, an error
message is emitted. If this recurs, inform Service.
FUNC 6
Self-Test
INJECT = Complete
CLEAN = Leak only
Turn pi to
select function
FUNC 6
Self-Test
Ps leak
_____
_____ hPa
Pw leak
_____
_____ hPa
Injection _____
_____
hPa
ps=7000
pw=
18 hPa
FUNC 6
Self-Test
Ps leak
_____
_____ hPa
Pw leak
_____
_____ hPa
ps=7000
pw=
18 hPa
5 Troubleshooting
Troubleshooting
5