5 troubleshooting, Troubleshooting – Eppendorf FemtoJet express User Manual
Page 35

83
Error 12
Supply pressure does not drop far enough
Possible cause: Technical defect
Solution: Service
Error 13
Supply pressure does not rise
Possible cause: Technical defect
Solution: Service
Error 14
Massive leak in pressure accumulator
Possible cause: Technical defect
Solution: Service
Error 15
Leak in pressure accumulator
Possible cause: Technical defect
Solution: Service
Error 16
Leak through Valve V1 is too large
Possible cause: Technical defect
Solution: Service
Error 20
Working pressure has not been reached
Compressor runs without supply pressure being reached
within a specific time.
Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.
Solution:
Check the capillary. Check the capillary holder and tube for
leaks or tighten up the tube connections.
Service.
ERROR 012
Supply pressure
does not fall
INJECT = Continue
ERROR 013
Supply pressure
does not raise
INJECT = Continue
ERROR 014
Supply pressure
leak is too high
INJECT = Continue
ERROR 015
Supply pressure
leak is too high
INJECT = Continue
ERROR 016
Valve 1 pressure
leak is too high
INJECT = Continue
ERROR 020
Working pressure
cannot be reached
INJECT = Continue
5 Troubleshooting
Troubleshooting
5
05_Fehlersuche_en.fm Seite 83 Freitag, 29. Juni 2007 2:50 14