5 constant pressure, 3 mode of operation 3, Pw= 50 hpa – Eppendorf FemtoJet express User Manual
Page 28: Mode of operation
76
3.7.4 Switching on/off the background illumination of the display
3.7.5 Constant pressure
FUNC 3
Use the
key to switch the illumination on or off. The
currently selected setting (in this case: ON) is underlined.
In the normal operating mode, illumination switches off after
three minutes if no operating control has been actuated.
When an operating control is actuated, the illumination
lights up again.
FUNC 4
This function can be used to apply a constant working
pressure pw in the range from 5 to 6,000 hPa (87.0 psi) at
the pressure outlet. The pressure unit psi can also be used.
The working pressure p
w
is set with the right-hand variable
regulator (p
c
), the specified value is shown in large
characters. The actual value is shown in the bottom right of
the display.
The Clean function can be activated. The pressure of the
Clean function is shown in the bottom right of the display in
the form of p
w
.
FUNC 3
Push Inject to select
Illumination
ON OFF
Turn pi to
select function
FUNC 4
Continuous flow
pw= 50 hPa
Supply pressure
5261
Turn pi to
pw[hPa]
select function
50
FUNC 4
Continuous flow
INJECT = start
Turn pi to
select function
Mode of operation
3 Mode of operation
3
03_Bedienung_en.fm Seite 76 Freitag, 29. Juni 2007 2:49 14