6 technical data, 6 technical data 6, Technical data – Eppendorf FemtoJet express User Manual
Page 40

88
Voltage/Frequency:
100 V – 240 V ± 10 %, 50 – 60 Hz
Set voltage:
adapts automatically
Fuse for 100 – 240 V:
T1.0 A, 250 V (2 pcs.)
Leistungsaufnahme:
20 W
Max. current consumption:
< 0.5 A
Protection class:
I
Overvoltage category:
II (IEC 61010-1)
Pollution degree:
2 (IEC 664)
Ambient temperature:
15 °C – 35 °C
Ambient relative humidity:
max. 70 %
Weight: 4.5
kg
Dimensions:
Width:
220 mm
Depth:
170 mm
Height: 280
mm
Minimum external pressure
to be supplied:
> 4,000 hPa ( 60 PSI), constant
Maximum external pressure
to be supplied:
< 8,000 hPa (120 PSI), constant
Injection time t
i
:
0.0 – 99.9 s
in 0.1 s increments
Injection pressure p
i
:
0; 5 – max. 6 000 hPa (87 PSI),
in 1 hPa (0.01 – 0.02 psi) increments
Compensation pressure p
c
:
0; 5 – max. 6 000 hPa (87 PSI),
in 1 hPa (0.01 – 0.02 psi) increments
Regulating accuracy:
1 % of value set in each case,
at least ± 1 hPa
Clean rinsing pressure:
max. approx. 7,000 hPa (102 psi)
RS 232 computer
Baud rate: 9,600
interface:
Start bit:
1
Data bits:
8
Parity: non
Stopbit:
2
Technical specifications subject to change!