An internal self-calibration of 3 seconds for more precise adjustment of the pressure takes
place every 10 minutes. No injection is carried out during this time.
Injection can continue as usual following the self-calibration.
An already activated injection will be carried out by the FemtoJet express. The self-calibration
takes place afterward.
Message during the self-calibration on the display of the FemtoJet express:

If the

FemtoJet express is connected with the InjectMan NI 2, the following message appears on

the display of the micromanipulator:

Self-calibration intervals:

– In the first hour after the machine is switched on: every 10 minutes.

– Afterwards:

Every 30 min at pc < 30 hPa or every 15 min at pc > 30 hPa
following the last activation.

A brief signal sounds at the beginning and end of the
calibration.

Press the joystick button again to repeat the injection.

WAIT

self calibration

Warning 001
Check Injector and cable
Press joystick button