5 troubleshooting, Troubleshooting – Eppendorf FemtoJet express User Manual

Page 37

background image

85

Error 26
Working pressure leak rate is too high

Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.

Solution:
Check capillary. Check capillary holder and tube for leaks or
tighten up tube connections.
Service.

Error 27
Working pressure leak rate is too high

Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.

Solution:
Check capillary. Check capillary holder and tube for leaks or
tighten up tube connections.
Service.

Error 28
Working pressure leak rate is too high

Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.

Solution:
Check capillary. Check capillary holder and tube for leaks or
tighten up tube connections.
Service.

Error 30
Sensor offset supply pressure is too low

Possible cause: Technical defect

Solution: Service