5 troubleshooting, Troubleshooting – Eppendorf FemtoJet express User Manual
Page 37

85
Error 26
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.
Solution:
Check capillary. Check capillary holder and tube for leaks or
tighten up tube connections.
Service.
Error 27
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.
Solution:
Check capillary. Check capillary holder and tube for leaks or
tighten up tube connections.
Service.
Error 28
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly attached.
(Broken) capillary with large opening.
Technical defect.
Solution:
Check capillary. Check capillary holder and tube for leaks or
tighten up tube connections.
Service.
Error 30
Sensor offset supply pressure is too low
Possible cause: Technical defect
Solution: Service
ERROR 026
Working pressure
leak is too high
INJECT = Continue
ERROR 027
Working pressure
leak is too high
INJECT = Continue
ERROR 028
Working pressure
leak is too high
INJECT = Continue
ERROR 030
Sensoroffset supply
is too low
INJECT = Continue