Maintenance
FemtoJet

®

4i

English (EN)

44

1. Wipe the painted parts and the aluminum surfaces with a cloth and mild detergent.

2. Polish with a dry cloth.

8.5

Hints with regard to service intervals

The display shows the following message:

Please contact local service soon

– Contact the authorized service.

Please contact local service now

– Have the service performed by the authorized

service.

8.6

Service and maintenance

Application-specific service and safety inspections are not required.

The Eppendorf AG service team is available to service and certify your device.

Service provisions:

• Service

• Operational qualification (OQ) according to manufacturer's specifications

• Electrical safety testing according to country-specific regulations

• Software update

Information on the services offered can be found on our webpage: www.eppendorf.com/
epservices.

8.7

Disinfection/decontamination

Prerequisites

• All device parts are cleaned.

• A disinfectant containing alcohol (such as isopropanol or spirit) is available.

Wipe all parts of the device with a cloth and the disinfectant.

Choose the disinfection method which corresponds to the legal regulations and
guidelines in place for your range of application.
If you have any questions on cleaning, disinfection, decontamination, or the
cleaning agents to be used, please contact Eppendorf AG Application Support.
The relevant contact details are provided on the back of this operating manual.