Troubleshooting
FemtoJet

®

4i

English (EN)

40

7.2.4

Error 39 – 40

Symptom/message

Cause

Remedy

Warning #39

• Capillary is broken.

Insert a new capillary.

• Injection tube is not

separated from the device
when switching on.

Take off the injection tube.

Wait until the initialization
phase is completed.

Connect the injection tube.

• Injection tube connected

without equipped capillary.

Connect an equipped
capillary.

• Storage pressure is too low.

Wait until the storage
pressure is reached.

• System error

Switch the device off and
back on again.

If the error persists, please
contact the authorized
service.

Warning #40

• Capillary is broken.

Insert a new capillary.

• Injection tube is not

separated from the device
when switching on.

Take off the injection tube.

Wait until the initialization
phase is completed.

Connect the injection tube.

• Injection tube connected

without equipped capillary.

Connect an equipped
capillary.

• Storage pressure is too low.

Wait until the storage
pressure is reached.

• System error

Switch the device off and
back on again.

If the error persists, please
contact the authorized
service.