4 error 39 – 40, Error 39 – 40 – Eppendorf FemtoJet 4x v.1 User Manual
Page 40

Troubleshooting
FemtoJet
®
4i
English (EN)
40
7.2.4
Error 39 – 40
Symptom/message
Cause
Remedy
Warning #39
• Capillary is broken.
Insert a new capillary.
• Injection tube is not
separated from the device
when switching on.
Take off the injection tube.
Wait until the initialization
phase is completed.
Connect the injection tube.
• Injection tube connected
without equipped capillary.
Connect an equipped
capillary.
• Storage pressure is too low.
Wait until the storage
pressure is reached.
• System error
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Warning #40
• Capillary is broken.
Insert a new capillary.
• Injection tube is not
separated from the device
when switching on.
Take off the injection tube.
Wait until the initialization
phase is completed.
Connect the injection tube.
• Injection tube connected
without equipped capillary.
Connect an equipped
capillary.
• Storage pressure is too low.
Wait until the storage
pressure is reached.
• System error
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.