6 operation, 1 switching on or off the microinjector, 1 switching on the microinjector – Eppendorf FemtoJet 4x v.1 User Manual

Page 26: 2 switching off the microinjector, 2 switching on or off the standby mode, 1 switching on the standby mode, Operation 6.1, Switching on or off the microinjector 6.1.1, Switching on the microinjector, Switching off the microinjector

background image

Operation
FemtoJet

®

4i

English (EN)

26

6

Operation

6.1

Switching on or off the Microinjector

6.1.1

Switching on the Microinjector

1. Take off the injection tube.

2. Switch on the Microinjector with the mains/power switch.
The Microinjector performs a self test.

The operating pressure is built up.

The main screen appears.

6.1.2

Switching off the Microinjector

1. Keep the

standby

key pressed.

The message

Hold Standby to exhaust

appears.

The pressure reservoir is emptied.

2. Switch off the Microinjector with the mains/power switch.

6.2

Switching on or off the standby mode

For short intermissions, you can use the standby mode. The Microinjector remains ready
for operation.

6.2.1

Switching on the standby mode

1. Press the

standby

key.

2. Take off the injection tube.
The display shows

STANDBY

.

The operating controls are deactivated.

The pressure in the pressure reservoir is maintained.

WARNING! Electric shock due to damage to device or mains cable.

Only switch on the device if the device and mains cable are undamaged.

Only use devices that have been properly installed or repaired.

In case of danger, disconnect the device from the mains supply by pulling the
power plug from the device or the mains socket or, by using the isolating
device intended for this purpose (e.g., emergency stop switch in the
laboratory).