FemtoJet

®

4i

English (EN)

31

Press the

inject

key.

The injection movement is triggered on the micromanipulator.

The

inject

key is disabled during the injection.

The elapsing injection time is displayed.

6.8.2

Manually injecting liquid

Prerequisites

• The injection parameters are set.

• The micromanipulator is connected.

• For the synchronization,

IMMEDIATE

is set on the micromanipulator.

1. Press the

auto man

key.

The display shows

Manual

.

2. Keep the

inject

key pressed.

The injection of the liquid continues until the

inject

key is released.

The progress of the injection time is displayed.

6.9

Replacing the capillary

1. Press the

menu enter

key.

2. Select the

Change capillary

menu.

3. Confirm with the

menu enter

key.

The message

Capillary may be changed now

is displayed.

The valve on the injection tube is closed.

4. Replace the capillary on the micromanipulator.

5. Press the

menu enter

key.

The main screen appears.

6.10

Rinsing the capillary

You can rinse a clogged capillary using the cleaning function.

1. Keep the

clean

key pressed.

The rinsing of the capillary continues until you release the

clean

key.

The capillary is rinsed at maximum pressure.
The pressure build-up is shown in a chart.

6.11

Setting the injection counter to zero

For each injection performed, the counter is increased by one. You can reset the counter
to zero.

1. Press the

count

key.

The

Count

display is reset to zero.