2 manually injecting liquid, 9 replacing the capillary, 10 rinsing the capillary – Eppendorf FemtoJet 4x v.1 User Manual
Page 31: 11 setting the injection counter to zero, Manually injecting liquid, Replacing the capillary, Rinsing the capillary, Setting the injection counter to zero
Operation
FemtoJet
®
4i
English (EN)
31
Press the
inject
key.
The injection movement is triggered on the micromanipulator.
The
inject
key is disabled during the injection.
The elapsing injection time is displayed.
6.8.2
Manually injecting liquid
Prerequisites
• The injection parameters are set.
• The micromanipulator is connected.
• For the synchronization,
IMMEDIATE
is set on the micromanipulator.
1. Press the
auto man
key.
The display shows
Manual
.
2. Keep the
inject
key pressed.
The injection of the liquid continues until the
inject
key is released.
The progress of the injection time is displayed.
6.9
Replacing the capillary
1. Press the
menu enter
key.
2. Select the
Change capillary
menu.
3. Confirm with the
menu enter
key.
The message
Capillary may be changed now
is displayed.
The valve on the injection tube is closed.
4. Replace the capillary on the micromanipulator.
5. Press the
menu enter
key.
The main screen appears.
6.10
Rinsing the capillary
You can rinse a clogged capillary using the cleaning function.
1. Keep the
clean
key pressed.
The rinsing of the capillary continues until you release the
clean
key.
The capillary is rinsed at maximum pressure.
The pressure build-up is shown in a chart.
6.11
Setting the injection counter to zero
For each injection performed, the counter is increased by one. You can reset the counter
to zero.
1. Press the
count
key.
The
Count
display is reset to zero.