Operation

FemtoJet

®

4i

English (EN)

27

6.2.2

Switching off the standby mode

Prerequisites

• The display shows

STANDBY

.

1. Take off the injection tube.

2. Press the

standby

key.

The Microinjector performs a short self test.

The display shows the main screen.

6.3

Determining the injection parameters

To determine the correct injection parameters, you can carry out a test injection with
fluorescence dye.

6.3.1

Filling the capillaries with fluorescence dye

Prerequisites

• Use a capillary with an opening of 0.5 μm.

• The injection tube is mounted to the universal capillary holder.

• Pipette with Microloader by Eppendorf is available.

1. Fill the Microloader with fluorescing liquid.

2. Equip the capillary with the Microloader.

3. Insert the capillary into the universal capillary holder.

6.3.2

Carrying out a test injection

Prerequisites

• The Microinjector and the micromanipulator are connected and ready for operation.

• The universal capillary holder is prepared with a capillary and fluorescence dye.

• The universal capillary holder is clamped in the Eppendorf micromanipulator.

• The Petri dish with adherent cells is prepared.

1. Connect the injection tube to the Microinjector.

2. Define the lower safety limit on the micromanipulator.

3. Position the capillary above the cell.

4. Press the

prog 1

key.

5. Press the

inject

key.

6. Check the injection visually.