Eppendorf FemtoJet 4x v.1 User Manual
Page 38
Troubleshooting
FemtoJet
®
4i
English (EN)
38
Error #13
• The compressor does not
work or is faulty.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #14
• Pressure control is faulty.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #15
• System error
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #16
• System error
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #17
• An open injection tube is
connected.
Take off the injection tube.
Connect the injection tube to
the universal capillary
holder.
Connect the injection tube.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #18
• An open injection tube is
connected.
Take off the injection tube.
Connect the injection tube to
the universal capillary
holder.
Connect the injection tube.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Symptom/message
Cause
Remedy