Troubleshooting
FemtoJet
®
4i
English (EN)
36
7
Troubleshooting
7.1
General errors
7.2
Error messages
7.2.1
Error 01 – 10
Symptom/message
Cause
Remedy
No liquid is injected.
• Capillary is clogged.
Clean the capillary using the
clean
function.
If the error persists, replace
the capillary.
Capillary does not
reach the injection
position.
• The lower safety limit is set
too high.
Set the lower safety limit on
the micromanipulator.
Symptom/message
Cause
Remedy
Error #01
• Software problem
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #02
• Compressor control signals
an unexpected error.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #03
• System error
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #04
• General electronic error in
the device.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.
Error #05
• Faulty hand or foot control.
Switch the device off and
back on again.
If the error persists, please
contact the authorized
service.