9. SYSTEM CALIBRATION

9.1 Calibration

The CS5467 provides DC offset and gain calibration
that can be applied to the voltage and current measure-
ments, and AC offset calibration which can be applied to
the voltage and current RMS calculations.

Since the voltage and current channels have indepen-
dent offset and gain registers, offset and gain calibra-
tion can be performed on any channel independently.

The data flow of the calibration is shown in

Figure 10

.

The CS5467 must be operating in its active state and
ready to accept valid commands. Refer to

7.6 Com-

mands

on page 25.

The value in the Cycle Count register (N) determines
the number of output word rate (OWR) samples that are
averaged during a calibration. DC offset and gain cali-
brations take at least N + T

SETTLE

samples. AC offset

calibrations take at least 6(N) + T

SETTLE

samples. As N

is increased, the accuracy of calibration results tends to
also increase.

The DRDY bit in the Status register will be set at the
completion of Calibration commands. If an overflow oc-
curs during calibration, other Status register bits may be
set as well.

9.1.1 Offset Calibration

During offset calibrations, no line voltage or current
should be applied to the meter. A zero-volt differential
signal can also be applied to the voltage inputs VIN1

VIN2 or current inputs IIN1 (IIN2of the CS5467.
(see

Figure 11

.)

9.1.1.1 DC Offset Calibration

The DC Offset Calibration command measures and av-
erages DC values read on specified voltage or current
channels at zero input and stores the inverse result in
the associated offset registers. This will be added to in-
stantaneous measurements in subsequent conver-
sions, removing the offset.

Gain registers for channels being calibrated should be
set to 1.0 prior to performing DC offset calibration.

9.1.1.2 AC Offset Calibration

The AC Offset Calibration command measures the re-
sidual RMS values read on specified voltage or current
channels at zero input and stores the inverse result in
the associated AC offset registers. This will be added to
RMS measurements in subsequent conversions, re-
moving the offset.

AC offset registers for channels being calibrated should
first be cleared prior to performing the calibration.

In

Modulator

+

X

V1, I1, V2, I2

Filter

N

I1

RMS

, V1

RMS

,

I2

RMS

, V2

RMS

I1

DCOFF

, V1

DCOFF

,

I2

DCOFF

, V2

DCOFF

I1

GAIN

, V1

GAIN

,

I2

GAIN

, V2

GAIN

0.6

+

= READABLE/WRITABLE REGISTERS.

N

+

X

N

1

DC

AVG

RMS

I1

ACOFF

, V1

ACOFF

,

I2

ACOFF

, V2

ACOFF

N

+

DC Gain

DC Offset

AC Offset

RMS

AC Gain

Negate

DC AVG

Negate

Figure 10. Calibration Data Flow

+

-

XGAIN

+

-

External
Connections

0V

+

-

AIN+

AIN-

CM +

-

Figure 11. System Calibration of Offset