Troubleshooting, Self test – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual
Page 64
Troubleshooting
1000
1010
Pressure sensor is defective Send the instrument out for repair.
1020
Pump running continuously
Send the instrument out for repair.
1040
1050
Pumping down takes too
long
Check whether the caps are firmly attached, and
that the adapter is firmly screwed in.
If they already are, then the pump is defective.
Send the instrument out for repair.
1030
1060
1070
1080
1090
Defective valve
Send the instrument out for repair.
1020
Aeration through the aera-
tion channel takes too long
Send the instrument out for repair.
1150
1160
Instrument leaks
Check the caps, adapter seating, and adapter filter,
and replace if necessary. Repeat the self-test. If
the error is not resolved, send the instrument out
for repair.
1170
Aeration through the
adapter takes too long
Remove the cap if this hasn't been done. If the
adapter filter is contaminated, replace the filter.
1200
Electronic defect
Send the instrument out for repair.
Error number
Cause
Corrective action
Self test
error: 1XXX
see instruction manual
back
restart
Error during the self-test
Error display
64