Troubleshooting, Self test – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual

Page 64

background image

Troubleshooting

1000
1010

Pressure sensor is defective Send the instrument out for repair.

1020

Pump running continuously

Send the instrument out for repair.

1040
1050

Pumping down takes too
long

Check whether the caps are firmly attached, and
that the adapter is firmly screwed in.
If they already are, then the pump is defective.
Send the instrument out for repair.

1030
1060
1070
1080
1090

Defective valve

Send the instrument out for repair.

1020

Aeration through the aera-
tion channel takes too long

Send the instrument out for repair.

1150
1160

Instrument leaks

Check the caps, adapter seating, and adapter filter,
and replace if necessary. Repeat the self-test. If
the error is not resolved, send the instrument out
for repair.

1170

Aeration through the
adapter takes too long

Remove the cap if this hasn't been done. If the
adapter filter is contaminated, replace the filter.

1200

Electronic defect

Send the instrument out for repair.

Error number

Cause

Corrective action

Self test

error: 1XXX

see instruction manual

back

restart

Error during the self-test

Error display