Troubleshooting, Error – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual

Page 63

background image

E

ng

lis

h

Test interruption

No pipette in adapter,
or not inserted firmly
enough

Insert pipette firmly
enough

Tested pipette is com-
pletely leaky

Pipette is defective, and
cannot be tested with this
testing instrument. Send
the pipette out for repair.

Tip does not seal onto
shaft

Insert another tip Check
shaft for damage

Wrong pipette type
adjusted

Select correct pipette
type in the menu.

Pipette adapter not
leak-tight

Check adapter for lint
or abrasion, and check
filter for moisture. Clean
adapter and exchange
filter. Perform self-test.

Pipette adapter is loose Tighten the lock nut

firmly. Perform self-test

Moisture in the mea-
surement room, adapter
filter is clogged.

Exchange adapter filter,
perform self-test. (Quick
Test)

with 5 ml and 10 ml
pipettes: Measurement
pressure too strong
for the piston spring.
Piston drawn downward
when pumping down.

Set the pipette type to
'5ml soft' or '10 ml soft'.

Measure leak rate
implausible

Moisture or leaks in the
PLT unit

Perform self-test (Full
Test).

PLT doesn't react at all,
or behaves unusually.

Software error, electri-
cal malfunction

Disconnect all cables
from the instrument. Re-
connect the power cable
after approx. 5 seconds.

Troubleshooting

Problem

Error display

Possible cause

Corrective action

Error

Test cancelled: Leaka-

ge rate not measu-

rable!

back

Errors during pipette tests

If the error cannot be resolved by the measures described, send the instrument out for repair.

Note: