Test preparation – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual
Page 44

Test preparation
Before	the	test,	one	must	decide	whether	a	dynamic	or	static	test	is	to	be	
performed.
With	the	dynamic test,	in	addition	to	the	detection	of	static	leaks,	one	can	
also	determine	whether	a	defective	piston	(contamination,	scratches)	has	
caused	a	leak.
During	the	measurement	time,	the	pipette	button	must	be	pressed	down	
completely	and	evenly	2-3	times.	The	associated	piston	movement	allows	
errors	to	be	recognized.
The	button	stroke	is	accomplished	relatively	slowly	with	5	ml	and	10	ml	
pipettes.
In	the	dynamic	testing	of	pipettes,	the	piston	should	be	in	the	upper	position	
at	the	beginning,	and	should	be	brought	back	to	that	position	at	the	end.
In	the static test,	by	contrast,	the	pipette	button	is	not	pressed	during	the	
test	procedure,	i.e.	the	piston	doesn't	move.	In	power	pipettes,	one	can	posi-
tion	the	piston	in	the	uppermost	location	or	at	the	1st	stop.
A	leak	that	occurs	during	the	piston	stroke	can	thus	not	be	determined	at	this	
point.
We recommend that the dynamic test be performed.
Dynamic or static test?
The test can take place with or without the pipette tip.
We	recommend	that	the	test	be	performed	with	a	mounted,	unused	tip	so	
that	the	entire	pipette	system	is	tested.
When	a	leak	has	been	identified,	the	test	can	be	repeated	without	a	tip	to	
determine	whether	the	leak	arises	from	the	tip	cone/tip	coupling	region.	
The	adapter	must	be	changed	(see	page	46)	to	be	able	to	perform	the	test	
without	the	tip.
The pipette should be held firmly during the test.
Test with or without pipette tip?
44
