Testing multichannel pipettes, Main menu, Channels – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual
Page 50: Nominal volume

The 'Main Menu' is called
up by pressing or turning the
select knob.
Set to 'Pipette Type' by
turning the select knob.
Press the 'Execute' button.
The 'Channels' submenu is
called up.
1.
Calling up the
Main Menu
2.
Selecting the
pipette type
3.
Confirming the
pipette type
Set to 'Multichannel' by
turning the select knob.
Press the 'Execute' button.
The 'Nominal Volume'
submenu is called up.
4.
Selecting the
multi-channel
pipette
5.
Confirming the
multichannel
pipette
Main Menu
Pipette Type
Self Test
Switch OFF
Settings..
.
select
back
Main Menu
Pipette Type
Self Test
Switch OFF
Settings..
.
select
back
Channels
Single Channel
Multichannel
select
back
Channels
Single Channel
Multichannel
select
back
Nominal Volume
50 µl
100 µl
200 µl
250 µl
select
back
'200 μl' is selected by turn-
ing the select knob.
6.
Selecting the
pipette volume
Nominal Volume
50 µl
100 µl
200 µl
250 µl
select
back
Testing multichannel pipettes
Example: 200 μl 8-channel air-displacement pipette, dynamic test
Before testing, mount a new, unused, disposable tip, and insert four channels with tips in the previously
selected and attached Adapter 3 (see page 46).
What to do?
How to do it
Keys to press
Display readout
50