Testing multichannel pipettes, Ready, Test – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual

Page 51: Pass, Fail

background image

E

ng

lis

h

ready

MC – 200 µl

start

Test

preparing...

restart

back

Test

-121.21 hPa
restart

back

4s

(pressure value is only an
example)

PASS

start

MC – 200 µl

Q

L

: 0.04 hPa

*

ml/s

After the test is finished, the
display indicates the leak
rate that was determined and
whether the system passed
the test.

The 'Start/Execute' button
can be used to repeat the
test, or the test can be
continued with the next 4
channels.
Turning the select knob leads
directly to the Main Menu.

FAIL

start

MC – 200 µl

Q

L

: 4.6 hPa

*

ml/s

9.

Ending the test

Testing multichannel pipettes

green

red

The missing volume can be approximately determined
using the correlation table on page 61.

Press the 'Start/Execute'
button. The 'Start' submenu
is called up.

Press the 'Start/Execute'
button.
The test is begun, and
the test pressure is set (in
progress...)

Shortly before the end of the
preparation period, the LED
light shines 1 time red and 1
time green, then the progress
bar appears and the sys-
tem automatically starts the
measurement. At this time,
the pipette button is pressed
down evenly 2-3 times. The
piston must return the initial
position before the measure-
ment is finished! The prog-
ress bar graphic indicates the
remaining time and the cur-
rently achieved vacuum.
The LED indicator blinks or-
ange.

7.

Confirming the

pipette volume

8.

Starting the test

Press the pipette

button 2-3 times

orange

(leak rate is only an example)

(leak rate is only an example)

What to do?

How to do it

Keys to press

Display readout

Note:

51