Perform self test full test, Main menu, Self test – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual
Page 42

–
The self test determines if there are any leaks in the BRAND PLT unit. Two different tests are available.
The Quick Test lasts only about 2 minutes, and should be performed on a daily basis. The Full Test lasts
about 12 minutes. It includes the additional differentiated testing of the internal tanks, and should be
performed on a weekly basis.
Perform self test
Full test
The 'Main Menu' is called
up by pressing or turning the
select knob.
Set to 'Self Test' by turning
the select knob.
Press the 'Execute' button.
The 'Self Test' submenu
is called up. The last test
to be performed appears
enframed.
1.
Calling up the
Main Menu
2.
Setting up the
self-test
3.
Confirming the
self-test
Set to 'Selecting the Full
Test' by turning the select
knob.
Press the 'Start/Execute'
button.
The 'Start Menu' submenu
is called up.
Press the provided plug
firmly into the adapter (for
the multi-channel adapter,
seal all 4 channels).
4.
Selecting the
Full Test
5.
Confirming the
Full Test
6.
Seal the system
Main Menu
Pipette Type
Self Test
Switch OFF
Settings..
.
select
back
Main Menu
Pipette Type
Self Test
Switch OFF
Settings..
.
select
back
Self Test
Quick Test
Full Test
select
back
Self Test
Please insert
blind plug
start
back
Self Test
Quick Test
Full Test
select
back
What to do?
How to do it
Keys to press
Display readout
42