Testing single-channel pipettes

The 'Main Menu' is called
up by pressing or turning the
select knob.

Set to 'Pipette Type' by
turning the select knob.

Press the 'Execute' button.
The 'Channels' submenu is
called up.

1.

Calling up the

Main Menu

2.

Selecting the

pipette type

3.

Confirming the

pipette type

Set to 'Single-Channel' by
turning the select knob.

Press the 'Execute' button.
The 'Nominal Volume'
submenu is called up.

4.

Selecting the

single-channel

pipette

5.

Confirming the

single-channel

pipette

Main Menu

Pipette Type

Self Test

Switch OFF

Settings..

.

select

back

Main Menu

Pipette Type

Self Test

Switch OFF

Settings..

.

select

back

Channels

Single Channel

Multichannel

select

back

Example: 100 μl Air-displacement pipette, dynamic test
Before testing, mount a new, unused, disposable tip, and insert the pipette with the tip in the previously
selected and attached Adapter 1 (see page 46).

Channels

Single Channel

Multichannel

select

back

Nominal Volume

25 µl

50 µl

100 µl

200 µl

select

back

'100 μl' is selected by turn-
ing the select knob.

6.

Selecting the

pipette volume

Nominal Volume

25 µl

50 µl

100 µl

200 µl

select

back

Continued on next page

What to do?

How to do it

Keys to press

Display readout

47