beautypg.com

Correlation table, Relation- ship of q, Eng lis h – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual

Page 61

background image

E

ng

lis

h

Correlation table

Relation-
ship of Q

L

to

volume loss
(approx.)

Pipette tested

Leak rate Q

L

Pipette volume
loss (approx.)

Channels

Nominal volume

[hPa*ml/s]

[μl]

1-channel

1 μl

1

0,02

1-channel

2 μl

2

0,035

1-channel

2,5 μl

2

0,04

1-channel

5 μl y

5

0,05

1-channel

5 μl

2

0,05

1-channel

10 μl y

6

0,1

1-channel

10 μl c/n

2

0,1

1-channel

20 μl y

7

0,16

1-channel

20 μl c

3

0,16

1-channel

25 μl

7

0,2

1-channel

50 μl

8

0,4

1-channel

100 μl

5

0,6

1-channel

200 μl

6

1,2

1-channel

250 μl

6

1,5

1-channel

300 μl

6

1,8

1-channel

500 μl

8

3

1-channel

1000 μl

10

6

1-channel

1250 μl

15

7,5

1-channel

1500 μl

32

9

1-channel

2000 μl

59

15

1-channel

2500 μl

39

15

1-channel

5 ml

31

30

1-channel

5 ml soft

15

30

1-channel

10 ml

38

60

1-channel

10 ml soft

19

60

4-channel

1 μl

1

0,02

4-channel

2 μl

2

0,035

4-channel

2,5 μl

2

0,04

4-channel

5 μl

2

0,05

4-channel

10 μl

2

0,1

4-channel

20 μl

3

0,16

4-channel

25 μl

7

0,2

4-channel

50 μl

8

0,4

4-channel

100 μl

5

0,6

4-channel

200 μl

6

1,2

4-channel

250 μl

6

1,5

4-channel

300 μl

6

1,8

4-channel

500 μl

11

3

4-channel

1000 μl

9

6

4-channel

1250 μl

12

7,5

With a leak of a given
magnitude, the volume
loss of the pipette dur-
ing pipetting depends
on various factors, e.g.,
the tip opening.
(Tolerance for the mea-
sured leak rate: ± 20%)

61