Correlation table, Relation- ship of q, Eng lis h – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual
Page 61

E
ng
lis
h
Correlation table
Relation-
ship of Q
L
to
volume loss
(approx.)
Pipette tested
Leak rate Q
L
Pipette volume
loss (approx.)
Channels
Nominal volume
[hPa*ml/s]
[μl]
1-channel
1 μl
1
0,02
1-channel
2 μl
2
0,035
1-channel
2,5 μl
2
0,04
1-channel
5 μl y
5
0,05
1-channel
5 μl
2
0,05
1-channel
10 μl y
6
0,1
1-channel
10 μl c/n
2
0,1
1-channel
20 μl y
7
0,16
1-channel
20 μl c
3
0,16
1-channel
25 μl
7
0,2
1-channel
50 μl
8
0,4
1-channel
100 μl
5
0,6
1-channel
200 μl
6
1,2
1-channel
250 μl
6
1,5
1-channel
300 μl
6
1,8
1-channel
500 μl
8
3
1-channel
1000 μl
10
6
1-channel
1250 μl
15
7,5
1-channel
1500 μl
32
9
1-channel
2000 μl
59
15
1-channel
2500 μl
39
15
1-channel
5 ml
31
30
1-channel
5 ml soft
15
30
1-channel
10 ml
38
60
1-channel
10 ml soft
19
60
4-channel
1 μl
1
0,02
4-channel
2 μl
2
0,035
4-channel
2,5 μl
2
0,04
4-channel
5 μl
2
0,05
4-channel
10 μl
2
0,1
4-channel
20 μl
3
0,16
4-channel
25 μl
7
0,2
4-channel
50 μl
8
0,4
4-channel
100 μl
5
0,6
4-channel
200 μl
6
1,2
4-channel
250 μl
6
1,5
4-channel
300 μl
6
1,8
4-channel
500 μl
11
3
4-channel
1000 μl
9
6
4-channel
1250 μl
12
7,5
With a leak of a given
magnitude, the volume
loss of the pipette dur-
ing pipetting depends
on various factors, e.g.,
the tip opening.
(Tolerance for the mea-
sured leak rate: ± 20%)
61