beautypg.com

Test preparation – BrandTech Scientific PLT unit Pipette Leak Testing Unit User Manual

Page 45

background image

E

ng

lis

h

To protect the instrument from moisture and contamination, a readily change-
able PE filter is included in the bottom the silicone adapter.

1

2

3

Changing the adapter

A total of 3 adapters are available

Adapter 1: already premounted,

for testing single-channel pipettes up to 10 ml
with a mounted pipette tip

Adapter 2: included in the delivery,

for testing single-channel pipettes without a pipette tip

Adapter 3: available separately (optional), for testing of multi-channel

pipettes with or without a pipette tip, wherein all 4 channels are
tested simultaneously. The 4-channel adapter can be used for
4-, 8- and 12-channel pipettes.

When a leak is detected, three channels can be closed with
plugs. The leaking channel can be determined through subse-
quent single-channel testing.

During multi-channel pipettes testing, the tips can be allowed
to remain in the adapter when the instrument is disconnected.
These can be removed simply by hand, or left inserted for the
next test.

Test preparation

Testing of single- or multi-channel pipettes

Important!

45