beautypg.com

MagTek InSpec 9000 User Manual

Page 77

background image

Appendix C. Card Standards

67


The ISO 7811-4, -5, and -6 parameters for cards that the InSpec 9000 Card Tester measures are
as follows:

TRACK 1

Parameter

ISO Term

New Card Limit

Used Card Limit


Signal Amplitude*

Ui

126%

≥ Ui ≥ 64%

126%

≥Ui ≥ 52%

Average Bit Density

Ba

± 8%

± 8%

Flux Transition Spacing Variation

Bin

± 10%

± 15%

Subinterval Spacing Variation

Sin

± 12%

± 20%

Adjacent Bit Cell Variation

Bin+1

± 10%

± 15%

Adjacent Subinterval Variation

Sin+1

± 12%

± 30%


Start Sentinel Position

0.293" ± 0.020" (7.44 ± 0.50 mm)
per ISO 7811-4


* as a percent of Reference Signal Amplitude

TRACK 2

Parameter

ISO Term

New Card Limit

Used Card Limit


Signal Amplitude*

Ui

126%

≥ Ui ≥ 64%

126%

≥ Ui ≥ 52%

Average Bit Density

Ba

± 5%

± 8%

Flux Transition Spacing Variation

Bin

± 7%

± 15%

Subinterval Spacing Variation

Sin

± 10%

± 20%

Adjacent Bit Cell Variation

Bin+1

± 10%

± 15%

Adjacent Subinterval Variation

Sin+1

± 12%

± 30%


Start Sentinel Position

0.293" ± 0.020" (7.44 ± 0.50 mm)
per ISO 7811-4


* as a percent of Reference Signal Amplitude