beautypg.com

Figure 4-2. numerical parameters – MagTek InSpec 9000 User Manual

Page 42

background image

InSpec 9000 Encoded Card Tester

32


These numbers are the actual values of the bar charts, and they are color coded on the screen
accordingly (simulated here by shading). As the cursor is placed on any parameter number, the
actual ISO limits will appear at the bottom of the screen. The examples at the bottom of the
illustration (Green Text 125 Yellow Text 135, and Red Text 141) are shown near the top of the
illustration under Track 1 as Amplitude Min, Amplitude Avg, and Amplitude Max respectively.

Figure 4-2. Numerical Parameters

If a print is required, click on the Print button, and the results will be formatted as shown in
Figure 4-3. The screen will appear with a place for a description; this may be filled in or left
blank. The date and time will be printed. The values of the card read and ISO limits for new
cards and returned cards will be listed along with a measurement uncertainty statement that gives
the parameters (in percentage) for the Tester. The lower half of Figure 4-3 shows all limits for
cards with the colors (at the top of the columns) indicating the range of the limits; for example,
the Amplitude for Track 1 shows Red-52-Yellow-64-Green-126-Yellow-136-Red. Green falls
within 64 and 126. Red is under 52 and over 136.