beautypg.com

3 module 3 - d, 1dsp-t, Yes sel – Red Lion PAXLA User Manual

Page 8: No zero, No scrol, 3d-lev, Dsp rst, Off b-lit, Isplay, Ront

background image

8

5.3 Module 3 - d

Isplay

and

F

ront

p

anel

B

utton

p

araMeters

(

3-dSP

)

PAR

Pro

Front Panel

Display

Select Enable

Display

Update Time

Front Panel

Reset Enable

Programming

Security Code

Zero Display

W/Display

Reset

3-dSP

dSP-t

SEL

rSt

ZErO

CodE

Display

Intensity

Level

d-LEV

ScroL

Display

Scroll

Enable

b-LIt

Units Label

Backlight

PARAMETER MENU

1

dSP-t

DISPLAY UPDATE TIME

This parameter sets the display update time in seconds.

1

0.5

2

seconds

The

yES

selection allows the

SEL

button to toggle through the enabled

displays.

FRONT PANEL DISPLAY SELECT ENABLE (SEL)

yES

SEL

yES

NO

This parameter enables the

RST

button or user input to zero the input display

value, causing the display reading to be offset.

Note: For this parameter to operate, the

RST

button or User Input being used

must be set to

dSP

and the Input value must be displayed. If these conditions are

not met, the display will not zero.

ZERO DISPLAY WITH DISPLAY RESET

NO

ZErO

yES

NO

The

yES

selection allows the display to automatically scroll through the

enabled displays. The scroll rate is every 4 seconds. This parameter only appears

when the MAX or MIN displays are enabled.

DISPLAY SCROLL ENABLE

NO

ScroL

yES

NO

Enter the desired Display Intensity Level (1-3). The display will actively dim

or brighten as levels are changed.

DISPLAY INTENSITY LEVEL

1

to

3

3

d-LEV

This selection allows the

RST

button to reset the selected value(s).

FRONT PANEL RESET ENABLE (RST)

dSP

rSt

HI-LO

LO

HI

NO

dSP

OFF

b-LIt

UNITS LABEL BACKLIGHT*

The Units Label Kit Accessory contains a sheet of custom unit overlays

which can be installed in to the meter’s bezel display assembly. The backlight

for these custom units is activated by this parameter.

ON

OFF

Current Calibration

1. Connect the negative lead of a precision DC current source with an accuracy

of 0.01% or better to the COMM terminal. Leave the positive lead of the DC

current source unconnected.

2. With the display at

CodE 48

, press the

PAR

button. Unit will display

CAL

NO

3. Press the

RST

button to select the range to be calibrated.

4. Press the

PAR

button. Display reads

0.0A

5. With the positive lead of the DC current source unconnected, press

PAR

.

Display reads

CALC

for about 8 seconds.

6. When the display reads the selected range, connect the positive lead of the

DC

current source to the current input and apply full-scale input signal for the

range. (Note: For 200 mA range, apply 100 mA as indicated on the display.)

Press

PAR

. Display reads

CALC

for about 8 seconds.

7. Repeat steps 3 through 6 for each input range to be calibrated. When display

reads

CAL NO

, press the

PAR

button to exit calibration.

Voltage Calibration

1. Connect a precision DC voltage source with an accuracy of 0.01% or better

to the volt input and COMM terminals of the PAXLA. Set the output of the

voltage source to zero.

2. With the display at

CodE 48

, press the

PAR

button. Unit will display

CAL NO

.

3. Press the

RST

button to select the range to be calibrated.

4. Press the

PAR

button. Display reads

0.0v

.

5. With the voltage source set to zero (or a dead short applied to the input), press

PAR

. Display reads

CALC

for about 8 seconds.

6. When the display reads the selected range, apply full-scale input signal for

the range. (Note: For 200V range, apply 100V as indicated on the display.)

Press

PAR

. Display reads

CALC

for about 8 seconds.

7. Repeat steps 3 through 6 for each input range to be calibrated. When display

reads

CAL NO

, press the

PAR

button to exit calibration