USER INPUT ASSIGNMENT

Select the value(s) to which the User Input Function is assigned. The User

Input Assignment only applies if a selection of reset, or display hold is selected

in the User Input Function menu.

MODE

DISPLAY

No Function

NO

DESCRIPTION

User Input disabled.

Program Mode Lock-out

P-Loc

Zero Input

(Edge triggered)

ZErO

Zero the Input Display value causing

Display Reading to be Offset.

d-SEL

d-LEV

d-HLd

rESEt

Resets the assigned value(s) to the current

input value.

HI-LO

HI

dSP

LO

See Programming Mode Access chart

(Module 3).

Setpoint 1 and 2 Reset

Setpoint 1 Reset
Setpoint 2 Reset

Reset both setpoint 1 and 2 outputs.

Resets setpoint 1 output.
Resets setpoint 2 output.

rSt12

rSt-1

rSt-2

USER INPUT FUNCTION

NO

USrIN

dSP

U-ASN

Display Select

(Edge Triggered)
Display Intensity Level

(Edge Triggered)

Display Hold

Reset (Edge triggered)

Advance once for each activation.

Increase intensity one level for each

activation.

Holds the assigned display, but all other

meter functions continue as long as

activated (maintained action).

LO

U-Act

USER INPUT ACTIVE LEVEL

Select whether the user input is configured as active low or active high.

LO

HI

Entering Code 66 will overwrite all user settings with

the factory settings. The meter will display

rESEt

and then

return to

CodE 00

. Press the

PAR

button to exit the

module.

MIN DISPLAY ENABLE

YES

NO

NO

LO-En

NO

FCS

2.0

LO-t

MIN CAPTURE DELAY TIME

When the Input Display is below the present MIN value for the entered delay

time, the meter will capture that display value as the new MIN reading. A delay

time helps to avoid false captures of sudden short spikes.

0.0

to

999.9

sec.

Select

yES

to perform any of the Factory Service Operations shown below.

FACTORY SERVICE OPERATIONS

yES

NO

Enables the Minimum Display Capture capability.

The PAXLA uses stored calibration values to provide

accurate measurements. Over time, the electrical

characteristics of the components inside the PAXLA will

slowly change with the result that the stored calibration

values no longer accurately define the input circuit. For

most applications, recalibration every 1 to 2 years should be sufficient.

Calibration of the PAXLA involves a calibration which should only be

performed by individuals experienced in calibrating electronic equipment. Allow

30 minute warm up before performing any calibration related procedure. The

following procedures should be performed at an ambient temperature of 15 to 35

°C (59 to 95 °F).

CAUTION: The accuracy of the calibration equipment will directly affect the

accuracy of the PAXLA.

CALIBRATION

66

CodE

RESTORE FACTORY DEFAULT SETTINGS

Entering Code 50 will display the version (x.x) of the

meter. The display then returns to CodE 00. Press the

PAR

button to exit the module.

50

CodE

VIEW MODEL AND VERSION DISPLAY

48

CodE

5.2 Module 2 - s

econdary

F

unctIon

p

araMeters

(

2-SEC

)

PAR

Max Capture

Delay Time

Max Display

Enable

Min Display

Enable

Access Code

For Service

Operations

Min Capture

Delay TIme

Factory

Service

Operations

2-SEC

HI-En

HI-t

LO-En

LO-t

FCS

CodE

Pro

PARAMETER MENU

2.0

HI-t

MAX CAPTURE DELAY TIME

When the Input Display is above the present MAX value for the entered

delay time, the meter will capture that display value as the new MAX reading.

A delay time helps to avoid false captures of sudden short spikes.

0.0

to

999.9

sec.

NO

HI-En

MAX DISPLAY ENABLE

YES

NO

Enables the Maximum Display Capture capability.