Enabling and Disabling Logger Channels
The
Logger Channels
option permits you to enable or disable
the channels on the currently-connected logger.
Note:
During this process the logger’s memory is
cleared. If the information on the logger is still of
value, transfer the data to a PC before enabling
or disabling channels.
To enable or disable logger channels:
1
Select
Logger | Channels
.
2
In the window that appears, select or deselect the
checkbox to enable or disable a channel.
Click
OK
. This initiates the process of clearing the log-
ger’s memory. To conserve logger memory it is a good
idea to disable unused channels.
Note:
If you are using a temperature and humidity log-
ger, remember that the humidity sensor requires
temperature values; therefore, you cannot dis-
able the temperature sensor channel if you are
using the humidity channel.
Configuring Channels to Work with Transducers
This section applies only to those using Vaisala Veriteq VL-
4000 and SP-4000 Series Data Loggers. Used with
transducers, you can configure VL-4000 and SP-4000 Series
loggers to display measurements in units other than
milliAmps or volts.
To configure VL-4000 and SP-4000 Series Data Loggers to work
with transducers:
1
In vLog, choose
Logger | Setup
( ).
2
From the Setup window, click
Setup
next to the channel
to configure. Different loggers offer different channel
types, each with different setup windows and slightly