Operating environment, Inputs and outputs – Vaisala HMP110 User Manual
Page 68

User's Guide ______________________________________________________________________
66 __________________________________________________________________ M211060EN-E
Operating Environment
Operating temperature range
HMP60
-40 ... +60 °C
HMP63
-40 ... +60 °C
HMP110
-40 ... +80 °C
HMP113
-40 ... +60 °C
HMP110T
-40 ... +80 °C
Operating humidity range
0 ... 100 %RH, non-condensing
Electromagnetic compatibility
HMP60, HMP110 and HMP110T
EN 61326-1: Electrical equipment for
measurement, control and laboratory use -
EMC requirements – for use in industrial
locations.
EN 55022 Class B: Information technology
equipment - Radio disturbance characteristics
- Limits and methods of measurement.
HMP63 and HMP113
EN 61326-1: Electrical equipment for
measurement, control and laboratory use -
EMC requirements – Basic immunity test
requirements.
EN 55022 Class B: Information technology
equipment - Radio disturbance characteristics
- Limits and methods of measurement.
Inputs and Outputs
Operating voltage
with 0...1 V / 2.5 V or RS-485
5 ... 28 V
with 0...5 V / 1 ... 5 V
8 ... 28 V
Current consumption
typical average
1 mA
max peak
5 mA
Start-up time
probes with analog output
4 s at operating voltage 13.5 ... 16.5 VDC
2 s at other valid operating voltages
probes with digital output
1 s
Outputs
analog output channels
2
analog output types
0 ... 1 VDC / 2.5 VDC / 5VDC, 1 ... 5 VDC
with loop power converter
4 ... 20 mA (separate module, compatible
with humidity accuracy only)
digital output (optional)
RS-485 two-wire half-duplex
External loads
0 ... 1
R
L
min 10 k
Ω
0 ... 2.5 / 0 ... 5 / 1 ... 5 V
R
L
min 50 k
Ω