Vaisala HMP110 User Manual
Page 60
User's Guide ______________________________________________________________________
58 __________________________________________________________________ M211060EN-E
3.
Start a terminal application and set the correct connection settings.
The default serial settings are 19200 8 N 1. Remember to check
which COM port the USB cable is using.
4.
Wait for 20 – 40 minutes for the temperature to stabilize.
5.
Start the terminal session and connect the USB cable to the probe.
If your probe is in the analog output mode, you need to press Enter
a few times to start it in RS-485 mode.
6.
Verify that the connection works by giving the ? command:
?
If the probe does not respond with device information:
- Disconnect the USB cable from the probe and retry. If your probe
is in the analog output mode, press Enter a few times immediately
after connecting the USB cable.
- Try the ?? command in case the probe is in POLL mode, open the
line using OPEN command if necessary.
- Check your serial line settings and cable connections.
7.
When your serial connection is working, use the L command to see
the current user adjustment parameters.
l
8.
Use the ERRS command to see that no errors are active:
errs
9.
Give the CT command, with the temperature value of the reference
as a parameter:
ct 23.5
OK
10. Check with the L command that the user adjustment parameters
have changed.
If you wish to remove the effects of T calibration (returning the T
measurement of the probe to the factory calibrated state), see
section Clear Adjustment of T Measurement on page 46.