View Software Version

Use the VERS command to display the software version of the probe.

VERS

Example:

vers
HMP63 / 1.0.4

Serial Line Output Commands

Start Measurement Output

Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line.

For HMP60 and HMP110, the output always includes readings for
temperature, RH and Td. For HMP110T, the output includes only
temperature.

R

Example (HMP60 and HMP110):

r
T= 22.6 'C RH= 22.8 %RH Td= 0.3 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C
T= 22.6 'C RH= 22.5 %RH Td= 0.2 'C

Example (HMP110T):

r
T= 22.6 'C
T= 22.6 'C
T= 22.6 'C

Outputting the results continues in intervals issued with the command
INTV. You can stop the output by entering the S command.

Stop Measurement Output

Use the S command or press the Esc key to stop the continuous
measurement output.

S