beautypg.com

Output the measurement message once, Configuring serial line operation, Set serial line settings – Vaisala HMP110 User Manual

Page 41

background image

Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 39

Output the Measurement Message Once

Use the SEND command to output the measurement values once. If the
probe is in POLL mode and the line is not open for commands, specify
the address of the probe to receive the measurement message.

SEND [aaa]

where

aaa =

Address of the probe, range 0 ... 255. Set with the ADDR
command.


Example (probe in STOP mode, no address needed):

send
T= 22.7 'C RH= 20.0 %RH Td= -1.5 'C

Example (probe in POLL mode, with address 10):

send 10
T= 22.8 'C RH= 20.1 %RH Td= -1.3 'C

Configuring Serial Line Operation

Set Serial Line Settings

Use the SERI command to show or set the serial line settings. The new
settings will be taken into use when the probe is reset or powered up.

SERI [b p d s]

where

b

=

baud rate
(300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600)

p

=

parity (n = none, e = even, o = odd)

d

=

data bits (7 or 8)

s

=

stop bits (1 or 2)

Example (shows default settings):

seri
Baud P D S : 19200 N 8 1

This manual is related to the following products: