beautypg.com

Performance (hmp110), Relative humidity, Temperature – Vaisala HMP110 User Manual

Page 66: Dewpoint

background image

User's Guide ______________________________________________________________________

64 __________________________________________________________________ M211060EN-E

Performance (HMP110)

Relative Humidity

Measurement range

0 ... 100 %RH

Accuracy (incl. non-linearity, hysteresis
and repeatability)
temperature range +0 ... +40 °C
0 ... 90 %RH

±1.7 %RH

90 ... 100 %RH

±2.5 %RH

temperature range -40 ... 0 °C
and +40 ... +80 °C
0 ... 90 %RH

±3.0 %RH

90 ... 100 %RH

±4.0 %RH

Factory calibration uncertainty (+20 °C)

±1.5 %RH

Humidity sensor

Vaisala HUMICAP® 180R

Stability

±2 %RH over 2 years

Temperature

Measurement range

-40 ... +80 °C

Accuracy over temperature range
0 ... +40 °C

±0.2 °C

-40 ... 0 °C, +40 ... +80 °C

±0.4 °C

Factory calibration uncertainty

±0.2 °C

Temperature sensor

Pt1000 RTD Class F0.1 IEC 60751

Dewpoint

Measurement range

-40 ... +80 °C

Accuracy (incl. non-linearity,
hysteresis and repeatability)

temperature range 0 ... +40 °C
when dewpoint depression < 15 °C

±1 °C

when dewpoint depression 15 ... 25 °C

±2 °C

temperature range -40 ... 0 °C
and +40 ... +80 °C when
dewpoint depression < 15 °C

±2 °C

This manual is related to the following products: