Performance (hmp110), Relative humidity, Temperature – Vaisala HMP110 User Manual
Page 66: Dewpoint

User's Guide ______________________________________________________________________
64 __________________________________________________________________ M211060EN-E
Performance (HMP110)
Relative Humidity
Measurement range
0 ... 100 %RH
Accuracy (incl. non-linearity, hysteresis
and repeatability)
temperature range +0 ... +40 °C
0 ... 90 %RH
±1.7 %RH
90 ... 100 %RH
±2.5 %RH
temperature range -40 ... 0 °C
and +40 ... +80 °C
0 ... 90 %RH
±3.0 %RH
90 ... 100 %RH
±4.0 %RH
Factory calibration uncertainty (+20 °C)
±1.5 %RH
Humidity sensor
Vaisala HUMICAP® 180R
Stability
±2 %RH over 2 years
Temperature
Measurement range
-40 ... +80 °C
Accuracy over temperature range
0 ... +40 °C
±0.2 °C
-40 ... 0 °C, +40 ... +80 °C
±0.4 °C
Factory calibration uncertainty
±0.2 °C
Temperature sensor
Pt1000 RTD Class F0.1 IEC 60751
Dewpoint
Measurement range
-40 ... +80 °C
Accuracy (incl. non-linearity,
hysteresis and repeatability)
temperature range 0 ... +40 °C
when dewpoint depression < 15 °C
±1 °C
when dewpoint depression 15 ... 25 °C
±2 °C
temperature range -40 ... 0 °C
and +40 ... +80 °C when
dewpoint depression < 15 °C
±2 °C