Device information and status, View device information, Table 7 – Vaisala HMP110 User Manual

Page 38: Additional commands for probes with rs-485 output

background image

User's Guide ______________________________________________________________________

36 __________________________________________________________________ M211060EN-E

Table 7

Additional Commands for Probes with RS-485 Output

Command

Description

??

Output information about the device in
POLL mode

ADDR [0 ... 255]

Set the probe address (for POLL mode)

CLOSE

Close the temporary connection (Back to
POLL mode)

OPEN [0 ... 255]

Open a temporary connection to a POLL
mode device

SERI [baud p d s]

User Port settings (Default: 19200 N 8 1)
baud: 300 ... 57600

SMODE
[STOP/RUN/POLL]

Set the serial interface mode

Device Information and Status

View Device Information

The ? command outputs a listing of device information.

?

Example (output from HMP63):

?
HMP63 / 1.0.4
Serial number : H3640004
Batch number : T0001109
Sensor number : H0000322
Sensor model : Intercap
Order code : A12A0A2B0
Cal. date : 20120907
Cal. info : VAISALA/HEL
Time : 00:21:05
Serial mode : ANALOG
Baud P D S : 19200 N 8 1
Output interval: 1 S
Serial delay : 30
Analog delay : 10 S
Address : 0
Filter : 1.000
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 1 V
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C

If the probe is in poll mode, but a connection has not been opened using
the OPEN command, issue the ?? command. For a description of the
serial interface modes, see section Set Serial Line Settings on page 39.

??