FUTEK IPM500 (D500) Digital Display User Manual
Page 43
![background image](/manuals/372290/43/background.png)
43
Press Menu
Select Key
Press Digit Select
Key
Press Value Select
Key
_0000U
Line feed
0
No line feed after carriage return
1
Line feed after carriage return
_0000U
Alarm data with readings
0
No alarm data
1
Alarm data with reading
_0000U
Control of data output
0
Continuous data output
1
Data output on ASCII command only
.
SEr 2
.
Serial Setup 2
_0000U
Meter address with Custom
ASCII protocol
Select
1
thru
F
for addresses 1 thru 15.
Select
0.
thru
F.
(with decimal point) for
addresses 16 thru 31.
0
0000U
Half or full duplex
0
Full duplex
1
Half duplex
00000U
Special start & stop char.
0
Standard continuous mode
1
Special start & stop characters
00000U
RTS mode
0
Normal RTS
1
Single transmission
.
SEr 3
.
Serial Setup 3
00000U
Termination characters
0
Only at end of all items
1
At end of each item
.
SEr 3
.
Serial Setup 3
(continued)
00000U
Data sent in continuous
mode
0
Reading
1
Peak
2
Valley
3
Reading + peak
4
Reading + valley
5
Reading + peak + valley
__000U
Modbus ASCII gap timeout
0
1 sec
1
3 sec
2
5 sec
3
10 sec
__000U
Serial protocol
0
Custom ASCII
1
Modbus RTU
2
Modbus ASCII
SEr 4
.
Serial Setup 4.
__000U
Parity
0
None, 2 or more stop bits
1
Odd, 1 or more stop bits
2
Even, 1 or more stop bits
_
Addr
Modbus Address.
Appears only if a
Modbus protocol is
selected.
__000 __000 __000
Select digit to flash.
__247
Select
0
through
9
for flashing digit.
Address range is 1 to 247.